Skip to Main content Skip to Navigation
Journal articles

Synchrotron X-ray diffraction experiments with a prototype hybrid pixel detector

Abstract : A prototype X-ray pixel area detector (XPAD3.1) has been used for X-ray diffraction experiments with synchrotron radiation. The characteristics of this detector are very attractive in terms of fast readout time, high dynamic range and high signal-to-noise ratio. The prototype XPAD3.1 enabled various diffraction experiments to be performed at different energies, sample-to-detector distances and detector angles with respect to the direct beam, yet it was necessary to perform corrections on the diffraction images according to the type of experiment. This paper is focused on calibration and correction procedures to obtain high-quality scientific results specifically developed in the context of three different experiments, namely mechanical characterization of nanostructured multilayers, elastic-plastic deformation of duplex steel and growth of carbon nanotubes.
Complete list of metadata

Cited literature [16 references]  Display  Hide  Download

https://hal.archives-ouvertes.fr/hal-01203672
Contributor : Compte de Service Administrateur Ensam <>
Submitted on : Wednesday, September 23, 2015 - 3:24:12 PM
Last modification on : Friday, March 5, 2021 - 3:03:45 PM
Long-term archiving on: : Tuesday, December 29, 2015 - 9:38:42 AM

File

PIMM - JAC - LE BOURLOT - 2012...
Files produced by the author(s)

Identifiers

Citation

Christophe Le Bourlot, P Landois, S Djaziri, P.-O Renault, Eric Le Bourhis, et al.. Synchrotron X-ray diffraction experiments with a prototype hybrid pixel detector. Journal of Applied Crystallography, International Union of Crystallography, 2012, 45, pp.38-47. ⟨10.1107/S0021889811049107⟩. ⟨hal-01203672⟩

Share

Metrics

Record views

699

Files downloads

1340